Hotwire thermostat: Using fine copper wire as integrated sensor and heater for temperature control

Conventional thermostats are based on separate temperature sensor and heater devices with means for feedback between them. But in some recent EDN design ideas (DIs) we’ve seen thermostat designs that meld the functions of sensor and heater into a single active device (usually FET or BJT). The ploy can make a better fit to applications where the intended thermal load is physically small or has some other quirk of geometry that makes it inconvenient to apply the classic separate sensor/heater schema. This DI (see the figure) follows the melded concept but takes it in a somewhat different direction by using fine gauge copper wire (e.g., 40 AWG polyurethane insulated) as an integrated temperature sensor and heater.

Here’s how it works.

Miniature thermostat utilizing the tempco and I2R heating of 40 AWG copper wire as a melded sensor/heater.

Wow the engineering world with your unique design: Design Ideas Submission Guide

The resistance and temperature coefficient of a standard 40 AWG copper wire at 25 oC are generally spec’d at 1.07 Ω/foot and +0.393%/oC, respectively. Therefore, L feet of 40 ga can be expected to have an approximate resistance at a given temperature T of:

R(L,T) = 1.07 L(1 + 0.00393(T – 25))                 (1)
R = 1.07 L + 0.00421 L T – 0.00421 L 25           (2)
T = (R – 1.07 L + 0.00421 L 25) / 0.00421 L      (3)
T = (R – 0.965 L) / 0.00421 L                            (4)

Equation 4 holds well from R/L = 0.965 Ω/ft at 0o up to 1.6 Ω /ft at 155o (the recommended upper temperature limit for solderable polyurethane wire insulation). 

Consider the implications for the use of fine copper wire as a combination temperature sensor and heater.

If a suitable length (between 5 and 15 feet) of wire is placed in a feedback loop driving current through it so as to dissipate enough I2R heating to raise and maintain a temperature that creates a preselected constant wire resistance, then said temperature, and the temperature of any thermal load thermally bonded to it, would likewise be constant! This is exactly what the circuit in the figure does.

Q1’s drain supplies heating; heating current I to the sensor/heater wire (please ignore for a moment the minor contribution from start-up resistor R2). The voltage induced between the terminals of the R wire resistance is then:

V = IR                         (5)

This causes the A1b, Q2 current source to output:

I2 = V/(R4 + R7) = IR/(R4 + R7)           (6)

Which induces a voltage at pin 2 of A1b:

V2 = I2(R5 + R6) = IR(R5 + R6)/(R4 + R7)           (7)

Meanwhile, Q1’s source current (also equal to I) sampling resistor R1 produces:

V3 = IR1                     (8)

FET control amplifer A1a forces FET gate voltage and thereby R drive current such that:

V2 = V3                                           (9)
IR(R5 + R6)/(R4 + R7) = R1I          (10)
R = R1(R4 + R7)/(R5 + R6)             (11)

Thus, heater current, and therefore wire resistance and temperature, are forced to equilibrium values set purely by the resistance ratios listed in Equation 11, with the resultant constant temperature given by Equation 4.

About Q3. The thermostat circuit is intended to be as flexible as possible in regard to wire gauge, length and associated sensor/heater R resistance. To accommodate R < 10 Ω and consequent possibility of potentially damaging peak I values, Q3 removes Q1 gate drive when necessary and limits I to a safe ~1.4 A.

Setup and calibration. In further pursuit of flexibility in accommodating sensor/heater wire length and initial R, this simple calibration procedure is suggested for whenever the wire is replaced.

Before first power up, allow sensor/heater to fully equilibrate to room temperature.
Set R4 and R5 fully CCW.
Push and hold the CAL NC pushbutton.
Turn the power on.
Slowly turn R4 clockwise until LED first flickers on.
Release CAL.

Done. R5 is now “reasonably well” calibrated for a CCW to CW span of zero to 130oC above room temp.

Thermal coupling of the chosen length of sensor/heater wire to the desired thermal load (e.g., thermostated circuit component, test tube, petri dish, etc.) can be done by winding a meander of wire around the load, and securing it with polyimide tape, RTV silicone, or a similar heat tolerant adhesive.

And about R2. Although not significant in the steady state function of the circuit, without R2 the thermostat might be vulnerable to a failure to start when first switched on and might simply sit looking stupid. Indefinitely. Don’t ask how I know this…

Stephen Woodward’s relationship with EDN’s DI column goes back quite a long way. Over 100 submissions have been accepted since his first contribution back in 1974.

 Related Content

∆Vbe differential thermometer needs no calibration
Fixing a fundamental flaw of self-sensing transistor thermostats
Electromechanical relays: an old-fashioned component solves modern problems
Low-cost precision adjustable current reference and application
∆Vbe differential thermometer needs no calibration

<!–
googletag.cmd.push(function() { googletag.display(‘div-gpt-ad-native’); });
–>

The post Hotwire thermostat: Using fine copper wire as integrated sensor and heater for temperature control appeared first on EDN.

14 February 2024
http://institutionofelectronics.ac.uk/wp-content/uploads/2022/12/IOE_LOGO.png 0 0 http://institutionofelectronics.ac.uk/wp-content/uploads/2022/12/IOE_LOGO.png 2024-02-14 13:39:142024-02-14 13:39:14Hotwire thermostat: Using fine copper wire as integrated sensor and heater for temperature control

Latest news

  • Secrets of Oscilloscope Time Measurements14 August 2026 - 13:44
  • Radon: Level detection, risk determination, and as-needed mitigation13 August 2026 - 13:16
  • TI a first mover in CAN XL transceivers13 August 2026 - 10:13
  • Four-channel USB-UART IC boosts server management13 August 2026 - 05:08
  • eFuse speeds overcurrent detection13 August 2026 - 05:08
  • Memory platform tackles AI bottlenecks13 August 2026 - 05:08
  • 6.5-kV SiC MOSFET reaches 8-kV blocking13 August 2026 - 05:08
  • Made by Google 2026: This limited silicon-supply situation really sucks13 August 2026 - 05:08
  • Cheap and cheerful LMC555 RC PWM pulse generator12 August 2026 - 13:56
  • Record high wafer shipments. Can fabs keep pace?12 August 2026 - 07:51
IOE LOGO 2

Become a member

click here

Become a member

click here

Become a subscriber

click here

Become a sponsor

click here

© Copyright - The Institution of Electronics | Website by WHD Solutions
  • Link to LinkedIn
  • Link to Facebook
  • Link to X
Link to: Indoor solar cells spur design-option reassessments Link to: Indoor solar cells spur design-option reassessments Indoor solar cells spur design-option reassessments Link to: 5G router streamlines industrial operations Link to: 5G router streamlines industrial operations 5G router streamlines industrial operations
Scroll to top Scroll to top Scroll to top